Safety precautions • 11
English
NOTICE!
Inconsistent measuring results
Before you electronically save measurement values determined us-
ing this device and use them further (e.g. in seca PC software or in
a hospital information system), make sure that the measurement
values are plausible.
If measurement values are transmitted to seca PC software or a
hospital information system, make sure prior to further use that the
measurement values are plausible and are assigned to the correct
patient.
NOTICE
Results not comparable to other devices
Results of bioelectric impedance measurements are not interchange-
able with measurements obtained from different manufacturers‘
devices. Performing follow-up measurements using devices other
than seca medical Body Composition Analyzers may produce incon-
sistent data and lead to misinterpretation of measurement results.
Make sure to perform follow-up measurements using seca medical
Body Composition Analyzers only.
Dealing with packaging material WARNING!
Risk of suffocation
Packaging material made of plastic foil (bags) is a choking hazard.
Keep packaging material out of reach of children.
If the original packaging material is no longer available, only use
plastic bags with safety holes to reduce the risk of suffocation. Use
recyclable materials if possible.
NOTE
Keep the original packing material for future use (e.g. returning for
maintenance service).
2.3 Handling batteries and rechargeable batteries
Warning!
Personal injury due to improper handling
Batteries and rechargeable batteries contain harmful substances
which may explode if not handled properly.
Do not try to recharge batteries.
Do not expose (rechargeable) batteries to heat.
Do not burn (rechargeable) batteries.
If acid is leaking out, avoid contact with the skin, eyes and mucous
membranes. Rinse affected areas with plenty of clean water and
seek medical help at once.
NOTICE!
Damage to device and malfunctions with improper handling
Only use the type of (rechargeable) battery specified in this docu-
ment.
When replacing (rechargeable) batteries, always replace a complete
set at a time.
Do not short-circuit (rechargeable) batteries.
If you do not use the device for a long period of time, remove the
batteries (incl. rechargeable batteries). This prevents acid from leak-
ing into the device.
If acid leaked into the device, discontinue use. Have the device
checked by an authorised seca service partner and repaired if
necessary.