42 •
6. HYGIENE TREATMENT
WARNING!
Electric shock
The device is not de-energized when the on/off key is pressed and the
display goes out. Use of fluids on the device may cause electric
shock.
Ensure that the device is switched off before performing any hygiene
treatment.
Disconnect the power supply connector before performing any hy-
giene treatment.
Before each hygiene treatment , take the rechargeable battery out
of the device (if present and removable).
Ensure that no fluids penetrate the device.
CAUTION!
Damage to device
Inappropriate detergents and disinfectants may damage the sensitive
surfaces of the device.
Use only disinfectants free of chlorine and alcohol which are explic-
itly suitable for acrylic sheet and other sensitive surfaces (active in-
gredient: quaternary ammonium compounds, for example).
Do not use aggressive or abrasive cleaning agents.
Do not use organic solvents (e.g. white spirit or petroleum spirit).
6.1 Cleaning
Use a soft cloth dampened with mild soapsuds to clean the surfaces of
the device.
6.2 Disinfecting
1. Check that your disinfectant is suitable for sensitive surfaces and acrylic
sheet.
2. Follow the instructions for use for the disinfectant.
3. Disinfect the device:
Moisten a soft cloth with disinfectant and wipe down the device
Comply with the intervals: see table
6.3 Sterilizing
This device may not be sterilised.
Cleaning
Disinfecting
Sterilizing
Prior to every measurement with
direct skin contact
Measuring mat, electrode cables
After every measurement with
direct skin contact
Measuring mat, electrode cables
If required Monitor, storage compartment