6 •
Administration of seca patient files Measured results are administered in seca patient files. seca patient files can
be created directly on the device. With administrator rights, seca patient files
can be edited directly on the device.
The device can be configured so that the seca patient files of the device are
automatically synchronized with the patient files of the
seca analytics 115 PC
software.
seca patient files and seca patient databases contain exclusively data neces-
sary for working with seca products or which were determined using seca
products.
The export and import functions of the
seca analytics 115 PC software can be
used for exchanging data with practice and hospital information systems.
Evaluating measured results Bioimpedance measurements are evaluated in graphical form based on sci-
entifically-validated formulas. In-house studies by seca generated predictive
formulas for determining the parameters total body water (TBW), extracellular
water (ECW), fat-free mass (FFM) and skeletal muscle mass (SMM) for arms,
legs, torso and the whole body. Further studies generated normal ranges for
the following parameters: bioelectric impedance vector analysis (BIVA), mass
indices (FMI, FFMI), phase angle(φ), fat mass (FM) and skeletal muscle mass
(SMM).
Data transmission and network
functions
The device can be incorporated in a PC network via a LAN interface or
via WiFi. The device can thus communicate with the databases of
seca analytics 115 PC software and use the seca directprint function of
the seca analytics 115 PC software.
The monitor and the measuring mat communicate with one another via WiFi.
If WiFi is unavailable, the infrared interface in the storage compartment of the
monitor is used.
Scales and length measuring devices from the
seca 360° wireless system can
transmit measured results to the device wirelessly.
Compatibility This device (software version 1.0) is compatible only with Version
1.4 build 381 or higher of the seca analytics 115 PC software. There is no
downward compatibility with older versions of the
seca analytics 115.
1.3 User qualification
Administration/network connection The device may only be set up and incorporated in a network by experienced
administrators or hospital technicians.
Measuring mode The device and the seca analytics 115 PC software may only be operated by
persons with sufficient specialist expertise.
➔ Administration/network connection
➔ Measuring mode