Safety precautions • 9
English
CAUTION!
Patient hazard, malfunction
Keep other electrical medical devices, e.g. high-frequency surgical
devices, a minimum distance of approx. 1 meter away to prevent in-
correct measurements or wireless transmission interference.
Keep HF devices such as cell phones a minimum distance of ap-
prox. 1 meter away to prevent incorrect measurements or wireless
transmission interference.
The actual transmission output of HF equipment may require mini-
mum distances of more than 1 meter. Details can be found at
www.seca.com.
Preventing electric shock WARNING!
Electric shock
Set up devices which can be operated with the electricity supply so
that the power supply socket is within easy reach and the power
supply can be disconnected quickly.
Ensure that your local power supply matches the details on the de-
vice.
Connect this device only to a power supply with a protective earth
facility.
Do not connect the device to a power supply network if there is any
uncertainty about whether the protective earth is functioning. In this
case, use the device exclusively in rechargeable battery mode.
Never touch the power supply cable with wet hands.
Do not use extension cables or power strips.
Make sure that cables are not pinched or damaged by sharp edges.
Make sure that cables do not come into contact with hot objects.
Do not operate the device at an altitude of more than 3000 m above
sea level.
Preventing injuries and infections WARNING!
Injury from falls
Ensure that the device is positioned firmly and level.
Route connecting cables (if present) in such a way that neither user
nor patient can trip over them.
WARNING!
Risk of infection
Hygienically reprocess the scale regularly as described in the re-
spective section in this document.
Make sure that the patient has no infectious diseases.
Make sure that the patient has no open wounds or infectious skin
alterations, which may come into contact with the device.