28 •
Logging out/switching user Press the Logout button.
You will be logged out.
The login window is displayed.
Another user can log in Logging in.
Automatic standby The device automatically switches to standby if there are no entries on the
device for 5 minutes. This has the following effects:
the LED of the ON/OFF button is green.
the touchscreen display goes out.
all entries are retained.
the measuring mat remains switched on.
after you switch back on, you have to log in again.
operation can be continued.
If no entries are made for 10 minutes, the device switches off:
the LED of the ON/OFF button goes out.
data which have not been saved are lost.
the measuring mat switches off.
after you switch back on, you have to log in again.
entries have to be made again.
Switching the monitor to standby
Briefly press the ON/OFF button of the monitor.
the LED of the ON/OFF button is green. The touchscreen display goes
out.
The monitor is on standby.
Switching off WARNING!
Electric shock
The monitor cannot be de-energized by pressing the ON/OFF button.
Always remove the power supply connector and remove the re-
chargeable battery (if present and intended for technical purposes)
if the device needs to be de-energized, e.g. for the hygiene treat-
ment.
1. Ensure that the measuring mat is correctly suspended in the storage
compartment of the monitor Suspending measuring mat in storage
compartment.
2. Keep the ON/OFF button of the monitor pressed until the touchscreen
display and the LED of the ON/OFF button go out.
The device is switched off.
The measuring mat switches off automatically.
Logout