Technical data • 47
English
10.TECHNICAL DATA
10.1 Monitor
Monitor
Measuring mat
Bioimpedance measurement
Evaluation modules
Monitor
Dimensions
•Depth
•Width
•Height
230 mm
252 mm
262 mm
Net weight approx. 2 kg
Ambient conditions, operation
Temperature
Air pressure
•Humidity
+10 °C to +40 °C (50 °F to 104 °F)
700 hPa - 1060 hPa
30 % - 80 %, no condensation
Ambient conditions, storage
Temperature
Air pressure
•Humidity
-10 °C to +60 °C (14 °F to 140 °F)
700 hPa - 1060 hPa
0 % - 95 %, no condensation
Ambient conditions, transport
Temperature
Air pressure
•Humidity
-10 °C to +60 °C (14 °F to 140 °F)
700 hPa - 1060 hPa
0 % - 95 %, no condensation
Setup location, maximum altitude above mean sea level 3000 m
Display type 7" Touchscreen display
Power supply
Mains voltage
Mains frequency
Power consumption
Internal power supply unit
100 V ~ - 240 V ~
50 Hz - 60 Hz
0.85 A
Mobile power supply
Range (measuring)
Lithium-ion rechargeable battery
approx.8 h
Power consumption
Standby (touchscreen display off, ON/OFF button green)
In operation (ON/OFF button white)
Operation (charging rechargeable battery of monitor and measuring
mat, ON/OFF button white)
<5W
<9W
<35W
Medical device in accordance with Directive 93/42/EEC Class IIa
EN 60601-1: Medical electrical device, type BF
Type of protection IP20
Operating mode Continuous duty
Interfaces
seca 360° wireless
USB 2.0 (max 500 mA)
LAN: Ethernet (10/100 Base-T)
WiFi
Infrared
Inductive charging of rechargeable battery
for measuring mat
Compatible printer
Microsoft
®
Windows
®
-compatible printer
via seca analytics 115 PC software