Loading...
MTP-200 Series Users Manual
Shineway Technologies, Inc. - 31 -
Information includes: Event list, measurement parameters, markers information (A/B
marker, marking points and etc.), trace parameters and trace information.
Information and function windows description:
Label
Function
Meas. Para.
Measurement parameters:
Mode, Wavelength, Range, Pulse Width, Averaging Time and etc.
Events
Add event, modify event, delete event and analyze event.
Markers
A/B switch, move, lock, loss and reflection analysis.
Trace Para.
Trace testing parameters
Trace Info.
Cable and fiber information viewing and editing.
Source
OTDR module built-in visible light source.
Tab. 7.1 Information and Function Windows Description
Note: For detail information, please refer to “Trace Analysis and Event Operation”.
7.8.1 Switch Information and Function Windows
Click the labels of information and function windows to switch.
7.8.2 Events info reference
Click {Events} information window changes into events information.
Click [ ] and [ ] in scroll bars to look over events information.
7.8.3 Reanalyze
When not satisfied with threshold, change threshold, then reanalyze the trace by this
function to get satisfactory trace.
Reanalyze the trace:
1. If possible, look over or reset threshold.
2. In {Events}, click [Analyze], system will reanalyze the trace.
3. When the analysis finished, events list will show the new results.
Reanalyzing trace may cause added events are deleted, events revised are resumed,
and events deleted come back etc.