MTP-200 Series User’s Manual
Shineway Technologies, Inc. - 34 -
and gets more ideal test result.
Correct parameters setting is the essential condition of optical fiber accurate measurement.
So, when test manually, must set parameters according to requirements.
9.1 Common using parameters for OTDR testing
Following testing parameters displaying on {Meas. Para.} option card on the OTDR surface,
users can set conveniently.
Distance range of the testing
The width of laser pulse launched to optical fiber by OTDR
Testing time, choose the proper testing time
Laser wavelength, testing laser wavelength selecting
Testing mode, testing mode selecting
Tab. 9.1 OTDR common use parameters sheet
9.2 OTDR test parameters Configuration
Testing parameters in this section should be set in “OTDR Setting“, click [Setup] on the
OTDR surface to enter.
These testing parameters include: Optical fiber general setting, analysis parameters setting
and etc.
IOR of optic fiber which affects the transmission speed of
laser. Need to set separately according to working
wavelength relatively
Also be called back scatter, which affects backward scatter
power of laser in fiber.
Also be called splice loss detecting threshold. Non reflection
event with insertion loss greater than this threshold will be
displayed on trace.
Also called reflectivity. Reflection events with insertion loss
greater than this threshold will be displayed on trace.
Also be called optical fiber end detecting threshold. The first
event with insertion loss greater than this threshold is
identified as the end of fiber, and all following events will be
ignored on trace.