MTP-200 Series Users Manual
Shineway Technologies, Inc. - 60 -
12 Manual Trace Analysis
User can use markers and zoom tools to measure splice loss, attenuation and reflection, or
use markers to locate event position and relative power level.
12.1 Use Markers
Click {Markers} label in OTDR main interface to activate markers.
12.2 Event Location and Loss Information
OTDR application automatically calculates event location and display the location in events
list. User can manually check event location and distance between events. Loss information
can be also displayed. The X-axis represents distance and Y-axis represents relative power.
Fig. 12.1 Event location and loss information
To acquire event location and loss information
1. Click {Markers} label in OTDR main interface.
2. Place marker A or B to at the beginning of the event, the marker label will show
distance.
3. Fig 12.1 shows the insertion loss of current marker and A-B loss.