Event Options:
In {Event} options, there are various parameters of events listing, including event type,
number, position, insert loss, reflection loss, attenuation and accumulation and etc.
When switch to {Event}, application program will display a marker on the tracing graph.
When some event in event list is selected, the marker will move to that event position and
indicate.
When open multi trace, event options only display current trace.
Markers Options:
When switch to {Markers} options, application program display two Markers A and B, or
include some gauge points. Particular cases depend on the function button in the button area.
Move these Markers along the trace to calculate loss, attenuation, reflection and etc.
Can use the buttons in Markers area to relocate all Markers and also can drag them in trace
display window.
11.5 Switch And Move Markers
There are many ways can switch and move Markers, users can choose one from them or use
them in combination way to increase efficiency.
➢ In trace displaying area, use stylus or mouse select A/B Markers or label above
Markers, then can press and drag A/B move in the trace service area to the target
position.
➢ In the buttons area on the bottom of surface, click [ ] or [ ] to switch
between A and B.
➢ After select Markers, continuously click [ ] or [ ] to move Markers.
➢ After select marked point, click [ ], [ ], towards left or right to move the
Markers.