MTP-200 Series Users Manual
Shineway Technologies, Inc. - 39 -
Configuration zone is -10 to -70dB
9.11 Non-reflection Threshold (Splice Loss Detecting
Threshold) Setting
This configuration has direct impact on the listing of insertion loss events. Only insertion
loss event which is greater than this threshold will be listed. The value configuration zone is
0.00 to 5.00dB
The default value is 0.20dB.
9.12 Reflection Threshold (Reflection Detecting
Threshold) Setting
This configuration has direct impact on reflection events listing. Only event reflection which
is greater than this threshold will be listed. The value configuration zone is -10dB to -65dB.
The default value is -52.00dB
9.13 End Threshold (Optical Fiber End Detecting
Threshold) Setting
This threshold is the end threshold of optic fiber. If the end threshold equals 3.0dB, then the
first event with insertion loss greater than 3dB will be considered as the end of the optic
fiber.
The value configuration zone is 0.0dB to 20.0dB.
The default value is 3.00dB.