MTP-200 Series User’s Manual
Shineway Technologies, Inc. - 85 -
2. Front panel charging indicator ( ) also indicated charging status (Please refer
to 1.4 Front Panel Indicators Introduction).
20.7 Battery Instruction
◆ New battery is not fully charged, please charge the battery before the first time use
which takes approximately 3 hours.
◆ Charging time is subject to the type of functional modules in use, environment and
etc.
◆ After the battery is used for some time, its capacity may not match charging status
indication (e.g. charging indicator indicates full but MTP-200 may power off due
to low battery), then battery calibration is necessary (Please refer to 20.9 Battery
Calibration).
◆ Please do not store battery in the environment where temperature is over 60℃
(140F).
◆ Use specified charger to charge the battery.
20.8 Charge Battery
Plug the charger to MTP-200 and power outlet, charging process automatically starts and
finishes after battery is fully charged.
20.9 Battery Calibration
After the battery is used for some time, the battery gauge may not be accurate, e.g. it
indicates battery has run out but the battery can still provide power supply, or battery power
is zero, but the battery power can last a long time. Then the battery should be re-calibrated.