MTP-200 Series User’s Manual
Shineway Technologies, Inc. - 59 -
➢ File name: The name of current trace.
➢ Date: Date and time of test.
➢ Wavelength: Laser wavelength used for test.
➢ Pulse width: Laser pulse width used for sampling.
➢ Averaging time: Continuous sampling time.
➢ Range: Distance used for test.
➢ Link length: The length of optical fiber link calculated after test.
➢ Link loss: The total loss of the link starting point to the link end.
➢ Link attenuation: Average loss of the total link, indicated as function of distance.
➢ Event quantity: Total events number got from current trace analysis which is
according to test parameters
The content displayed on the right:
IOR, back scatter, splice threshold, reflection threshold, end detecting threshold.
➢ IOR: Indicate the trace IOR (also called group coefficient). If modify this
parameter, it will adjust the trace distance test.
➢ Back scatter: Rayleigh back scatter parameter setting of the displayed trace.
➢ Non-reflection threshold: Check the current setting of small non reflection event
during the trace analyzing.
➢ Reflection threshold: Check the current setting of small reflection event during the
trace analyzing.
➢ End threshold: Check the current setting of the loss of important events(will affect
the transmission performance during the trace analyzing)
Trace parameters options only displays the selected trace even multi traces are opened.
Please switch to the trace if you want to display the parameters of other traces. The trace
selected is in red color.