Figure 112: Menu Tree Loading/saving parameters in SOPAS ET
8.8
Loading/saving the parameters after replacing the analyzer
Load the configuration of the old analyzer module to the new analyzer module after
replacement. This can be done easily using SOPAS ET.
1. Insert the SD card of the previous electronics in the new electronics.
2. Switch the device on.
3. Login as “Authorized User”.
4. Select menu Maintenance/Replace analyzer.
5. Click <Load customer-specific parameters> (This loads just the customer-specific
data from the SD card).
6. Wait until the device restarts.