Holding Register No. Measured variable
4018 [32 Bit DInt] Device status
4020 [32 Bit Float, scaled according to VDI] NO2
4022 [32 Bit DInt] Device status
4024 [32 Bit Float, scaled according to VDI] NOx
4026 [32 Bit DInt] Device status
4028 [32 Bit Float, scaled according to VDI] NH3
4030 [32 Bit DInt] Device status
4032 [32 Bit Float, scaled according to VDI] CO
4034 [32 Bit DInt] Device status
4036 [32 Bit Float, scaled according to VDI] CH4
4038 [32 Bit DInt] Device status
4040 [32 Bit Float, scaled according to VDI] O2
4042 [32 Bit DInt] Device status
4044 [32 Bit Float, scaled according to VDI] VOC
4046 [32 Bit DInt] Device status
4064 [32 Bit Float, scaled according to VDI] Number of active measuring point
4066 [32 Bit DInt] Device status
4068 [32 Bit Float, scaled according to VDI] Number of triggered measuring point
4070 [32 Bit DInt] Device status
4072 [32 Bit Float, scaled according to VDI] pH washing water
4074 [32 Bit DInt] Device status
4076 [32 Bit Float, scaled according to VDI] PAH washing water (oil content)
4078 [32 Bit DInt] Device status
4080 [32 Bit Float, scaled according to VDI] Clouding washing water
4082 [32 Bit DInt] Device status
Remark: The device status is added for each measured value for conformity reasons.
Device status format. The individual status has the same significance as the device sta‐
tus of Discrete Inputs 1000 et seq.
Bit No.
0 Malfunction
1 Maintenance
2 Maintenance request
3 Outside the specification
4 Test operation according to VDI
11.3 Error messages and possible causes
Current pending messages are shown on the device display.
A combined list of messages is included in SOPAS ET (see "MARSIC300 Technical
Data").
NOTE
The following Table only includes those messages with classification "X" that are impor‐
tant for information.
Messages not included in the following Table have no further significance for operation.
ANNEX 11
8017585/YXD9/V3-1/2017-05 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC300
93
Subject to change without notice