NOTE
Display of current device state data: Menu Diagnosis/System param.
C = Classification
F = Failure
M = Maintenance request
U = Uncertain
E = Extended
Trigger: System
Table 43: Error codes - System
Code Error text C Description Possible clearance
S001 Temperature too high F Measuring cell temperature too high When T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check plug-in connector
on electronics unit.
When plug OK: Replace cell.
Optic head temperature too high When T < 356 °C:
When housing temperature >= 55 °C: Check
housing fan.
When housing temperature < 55 °C: Replace
electronics unit.
When T >= 356 °C: Check plug-in connector
on electronics unit.
When plug OK: Replace Analyzer module.
Temperature of heating for a subassembly
too high
Check device documentation to clarify which
subassembly is affected.
When T < 356 °C: Replace electronics unit.
When T >= 356 °C: Check subassembly plug-
in connector.
When plug OK: Replace subassembly
LPMS01 (1/2 control) temperature too high When housing temperature < 55°C: Electron‐
ics unit fan functioning?
Yes: Replace electronics unit.
No: Replace Analyzer module.
When housing temperature >= 55 °C: Check
housing fan.
LPMS02 (power electronics) temperature too
high
When housing temperature < 55 °C: Replace
electronics unit.
When housing temperature >= 55 °C: Check
housing fan.
S002 Temperature too low F After x minutes Check system documentation to clarify which
subassembly is affected (heating circuit 1 ..7).
Temperature displayed < -30 °C: Pt100
short circuit: Replace subassembly
For heated sample gas line: Connect
reserve Pt100
For analyzer: Replace Analyzer module
For optic head: Replace Analyzer mod‐
ule
Reset on circuit breaker under electron‐
ics unit possible: Check all cables
involved for damage (see "Circuit break‐
ers", page 86)
Check all plugs are plugged correctly.
Reset not possible: Replace subassem‐
bly affected
11 ANNEX
94
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/YXD9/V3-1/2017-05 | SICK