Code Error text C Description Possible clearance
S004 Flow too low F Flow too low Sample gas flow and instrument air flow too
low: Replace cell
Sample gas flow too low and instrument air
flow OK: Gas sampling system defective
Instrument air flow too low and sample gas
flow OK: Check all hose connections.
When all hose connections OK: Replace valve
block.
S005 Pressure too high F Pressure too high Only instrument air pressure too high:
Check and set connected instrument air.
Set correct pressure on pressure
reducer unit.
Only sample gas pressure too high:
Set sample gas pressure within device
specification
Instrument air and sample gas pressure too
high:
Exhaust gas hose crimped/blocked
Counter-pressure in exhaust duct too
high
Check all hose connections
If this does not work:
Replace the pressure control module
Otherwise: Replace Analyzer module
S006 Pressure too low F Pressure too low Replace pressure control module.
S008 Chopper F 24 V power supply malfunction: Replace elec‐
tronics unit.
24 V power supply OK: Replace Analyzer mod‐
ule.
S009 Motor filter wheel 1 F Filter wheel motor does not detect reference
position
24 V power supply malfunction: Replace elec‐
tronics unit.
24 V power supply OK: Replace Analyzer mod‐
ule.
S010 Motor filter wheel 2
S011 Motor filter wheel 3
S012 Emitter F Emitter voltage incorrect: Replace electronics
unit.
Emitter voltage OK and emitter power incor‐
rect: Replace emitter.
S013 5 Volt power F Replace electronics unit.
S014 24 Volt power F Replace electronics unit.
S015 Detector signal F Replace Analyzer module.
S016 Ref. energy too low F If further error messages are pending: Clear
corresponding error.
If no further error messages are pending:
Replace cell.
S024 No active component F When all "active" checkmarks of all compo‐
nents are inactive
Check in SOPAS ET.
S025 Evaluation module failure F Evaluation module could not be started "Load backup": Menu Maintenance/Save
parameters.
If error remains: "Load default setting" .
If error remains: Replace electronics unit.
S026 Evaluation mod. file error F Files for Evaluation module not created