10.8 Interfaces and protocols
Table 26: Interfaces and protocols
Operation and interfaces
Operation Via LC-Display or SOPAS ET software, several operating
levels, password-protected
Display and input Black-and-white foiled screen with function buttons
Status LEDs
•
"Power"
•
"Malfunction”
•
"Maintenance request"
Analog outputs Optional
Digital inputs/outputs Optional
Data interface 1 x Ethernet (Modbus TCP/IP)
Profibus Optional
Profinet Optional
Remote maintenance SICK MPR (optional)
PC operation SOPAS ET via Ethernet
10.9 Power supply
Table 27: Power supply
Power supply
Supply voltage (preset) IT network (without neutral conductor, not grounded)
•
3~230 V, PE
•
3~208 V, PE
3∼ 230 V L/L/PE
L1
L2
L3
PE
L1-L2 230W
L2-L3 230W
L3-L1 230W
Figure 122: IT network switching
TN(S) network (with neutral conductor, grounded)
•
3~230 V, N, PE
•
3~115 V, N, PE
3∼ 230 V L/N/PE
L1
L2
L3
N
PE
L1-N 230W
L2-N 230W
L3-N 230W
L1-L2 400W
L2-L3 400W
L3-L1 400W
Figure 123: TN(S) network switching
Further variants optional (e.g.: 1~230 V, 2~208 V)
b
Refer to the system documentation provided for
the default value.
Frequency 50/60 Hz
Power consumption
•
Analyzer
•
Heated sample gas line
•
Gas sampling system
•
Heated probe tube
Power consumption
•
Approx. 1000 VA
•
Approx. 95 VA/m
•
Approx. 450 VA
•
Approx. 450 VA