The available programs are system-specific: see System documentation.
Parameter
6.3.1.1 Concentrations of test gases
Menu: Adjustment/Parameter/Concentrations
This menu displays the concentrations of the span gases and internal adjustment filter.
The concentrations of the test gases can be set.
Figure 68: Menu Concentrations
1
Enter the concen‐
tration of the test
medium
2
Only internal signif‐
icance
6.3.1.2 Adjustment factors
Menu: Adjustment/Parameter/Adjustment factors
This menu serves to set the correction factors (adjustment filter wheel optional).
The measured value is computed with both adjustment factors.
Factor "Internal adjust." is set automatically to "1.000" when factor "Span gas" is
changed.
Checkmark “Active”: This measuring component is computed.
Figure 69: Menu Adjustment factors
1
Display: Current factor
2
Input: New factor
3
Confirm: Use new factor
6.3.1.3 Start times
Menu: Adjustment/Parameter/Start times
This menu displays the start times of the "Cyclic triggers (CT1 .. CT16)“.
Figure 70: Menu Start times
1
Cyclic trigger name
2
Next start time
Deactivation via operator
panel or menu: Parameter/
Variables and functions/
Cyclic trigger (CTi)
PARAMETERIZATION 6
8017585/YXD9/V3-1/2017-05 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC300
51
Subject to change without notice