Component Measuring range
H
2
O 0 ... 40% by vol.
Table 20: Measuring point switchover
Measuring point switchover
Measuring point switchover Max. 2 measuring points (optional 8 measuring points)
Table 21: Measured value characteristics
Measured value characteristics
Measuring principle Photometric
Measuring precision < 2% of the respective full scale value
Detection limit < 2% of the respective full scale value
Sensitivity drift < 2% of the respective full scale value per week
Zero drift < 2% of the respective full scale value per week
Span drift < 2% of the respective full scale value per week
Setting time t
90
< 140 s, total measuring path as from probe extraction
10.4 Ambient conditions
Table 22: Ambient conditions - in operation
Ambient conditions in operation
Installation location Below deck
Ambient temperature +0 ... +45 °C
Relative humidity < 90% (without condensate)
Air pressure 900 ... 1100 hPa
Degree of protection IP 54
Table 23: Ambient conditions - in storage
Ambient conditions in storage
Ambient temperature -20 ... +70 °C
Relative humidity < 90% (without condensate)
10.5 Sample gas conditions
Table 24: Sample gas characteristics
Sample gas at the measuring point Characteristic
Process temperature 10 ... 550 °C
Sample gas temperature subassem‐
bly:
Sample gas probe
Sample gas line
Cell
Temperature:
Approx. 200 °C
Approx. 200 °C
Approx. 200 °C
Process pressure -20 ... +200 hPa relative
Dust load < 200 mg/m
3
10 TECHNICAL DATA
80
T E C H N I C A L I N F O R M A T I O N | MARSIC300 8017585/YXD9/V3-1/2017-05 | SICK