EasyManua.ls Logo

SICK MARSIC300

SICK MARSIC300
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Code Error text C Description Possible clearance
S097 Backup denied E
S098 AM E
S112 failure IO-Node E
S113 Check sum error F BCK (I/O) shows the transfer made from
Master to Slave (controller) had an incorrect
checksum and the Slave did not save the
data.
Check I/O module, cable damage
S114 Communication error F COM (I/O) communication error with an /I/O
module.
S115 High/low voltage F PF0 (I/O) signals the internal voltage monitor‐
ing of the 5 V and 24 V supply voltages
exceed or underflow a range.
Please contact SICK Customer Service
S116 Output without current F TOO (HC3X)
This Table contains solution proposals that can only be processed by specially trained
personnel.
ANNEX 11
8017585/YXD9/V3-1/2017-05 | SICK T E C H N I C A L I N F O R M A T I O N | MARSIC300
97
Subject to change without notice

Table of Contents

Other manuals for SICK MARSIC300

Related product manuals