Code Error text C Description Possible clearance
S097 Backup denied E
S098 AM E
S112 failure IO-Node E
S113 Check sum error F BCK (I/O) shows the transfer made from
Master to Slave (controller) had an incorrect
checksum and the Slave did not save the
data.
Check I/O module, cable damage
S114 Communication error F COM (I/O) communication error with an /I/O
module.
S115 High/low voltage F PF0 (I/O) signals the internal voltage monitor‐
ing of the 5 V and 24 V supply voltages
exceed or underflow a range.
Please contact SICK Customer Service
S116 Output without current F TOO (HC3X)
This Table contains solution proposals that can only be processed by specially trained
personnel.