9.3 Setting up edge measurement
Setting up edge measurement
Pr
er
equisites:
All conditions for optimal object position, required minimum distances, object
transmission and transmission differences are fulfilled.
The MLG-2 is visible in SOPAS and in “Online” mode.
1. If necessary, make settings in the “Blanking” menu.
2. If necessary, make settings in the “Material teach-in” menu.
3. Run “Installation assistant” wizard.
4. If necessary, make other settings in the “Material teach-in” menu. In this case, run
“Material teach-in” again.
Edge measurement is set up. The result is on the “Start” and “Enhanced Sensing”
pages.
Example
1 2
6
4
3
5
7
Figure 51: Example for edge measurement - display in “Start” tab, here 1 object with 2 edges
1
Current measured value of edge E1 in mm
2
Current measured value of edge E10 in mm
3
Display of the attenuation (opacity) in % over the measurement field width
4
Measured value and label of edge E10 (marked in blue)
5
Measured value and label of edge E1 (marked in blue)
6
Detected and measured object (grayscale, depending on attenuation)
7
Blanked areas of the measurement field
More information:
see "Plannin
g", page 25
see "Edge measurement and edge designation", page 30
see "Setting up MLG-2 WebChecker (installation assistant)", page 55
see "“Blanking” menu", page 96
see "“Material teach-in” menu", page 97
9.4 Parameterizing functions: Calculation of width and center positions
You can parameterize up to 10 functions. The following functions are possible:
Calc
ulate width between 2 edges.
Calculate distance between 2 edges
COMMISSIONING VIA SOPAS 9
8025190/2020-01-13 | SICK O P E R A T I N G I N S T R U C T I O N S | MLG-2 WebChecker
61
Subject to change without notice