CLINITEK Status+ Analyzer Operator’s Guide 117
System Configuration
8. Select the authorized operator options.
9. To return to the Select Ready screen, perform the following steps:
a. Select Done twice.
b. On the Authorized Operator-Operators list screen, select
Exit.
c. Select Done 3 times.
View, Edit, Print, and Delete Operators
You can perform the following tasks for the authorized operators:
• View the operators list
Authorized Operator
Option
Description
Screen 1 of 2
Patient test • Enabled (default) lets the
operator run patient tests.
• Disabled prevents the
operator from running
patient tests.
QC test • Enabled lets the operator run
QC tests.
• Disabled prevents the
operator from running QC
tests.
Screen 2 of 2
Recall results • Enabled lets the operator
recall results.
• Disabled prevents the
operator from recalling
results.
Instrument set up • Enabled lets the operator set
up the system.
• Disabled prevents the
operator from setting up the
system.