12.5 Online Calibration Options (P355 - P358)
BW500 and BW500/L
170 Operating Instructions, 12/2016, A5E33482052-AD
12.5.7
P365 Multispan
Note
Not available with BW500/L.
Select the span reference to be applied for determination of rate and totalization.
Entry:
1 = multispan 1 (MS1), for product or condition A
f
2 = multispan 2 (MS2), for product or condition B
3 = multispan 3 (MS3), for product or condition C
4 = multispan 4 (MS4), for product or condition D
5 = multispan 5 (MS5), for product or condition E
6 = multispan 6 (MS6), for product or condition F
7 = multispan 7 (MS7), for product or condition G
8 = multispan 8 (MS8), for product or condition H
Refer to Multispan and P270, Auxiliary Input Function (6).
12.5.8
P367 Direct Zero Entry
Directly enters the zero reference count.
Direct entry is intended for use when replacing software or hardware and it is not convenient to
perform an initial zero at that time (
f
= 0).
Refer to Recalibration (Page 67) for execution.
12.5.9
P368 Direct Span Entry
Directly enters the span reference count for the span selected, 1 to 8 (P368-01 to -08).
Note
BW500/L has one span entry possible, no indices will be presented.
Direct entry is intended for use when replacing software or hardware and it is not convenient to
perform an initial span at that time (
f
= 0).
Refer to Recalibration (Page 67) for execution.