12.12 SmartLinx Hardware Testing
BW500 and BW500/L
188 Operating Instructions, 12/2016, A5E33482052-AD
12.12.1
P790 Hardware Error
The results of ongoing hardware tests in the communications circuitry.
Values
Description
PASS
f
No errors
FAIL least significant byte (LSB) is given firstError
occurred communicating with card; device will try
to reinitialize communications with card. If mes-
sage continues, record values in P791 and P792
and contact your local Siemens representative.
ERR1 No module installed, or module not supported;
SmartLinx communications have been disabled
If FAIL or ERR1 is displayed in P790 (Hardware Error), go to P791 (Hardware Error Code) and
P792 (Hardware Error Count) for information about the error.
12.12.2
P791 Hardware Error Code
Indicates the precise cause of
Fail
or
ERR1
condition from P790.
Values
Description
0
f
No errors
Any other value Error code; provide this code to your Siemens
representative for troubleshooting
12.12.3
P792 Hardware Error Count
A count that increments by 1 each time Fail is reported in P790 (Hardware Error).
Values
Description
Range: 0 to 9999 Error count; provide this number to your Siemens
representative for troubleshooting.
12.12.4
P794 SmartLinx Module Type
This parameter is used to identify the module type when SmartLinx is used. If you are not using
SmartLinx, this parameter is not functional. Please see the associated SmartLinx instruction
manual for a full description of this parameter.