Start Up
6.3 Initial Start Up
BW500 and BW500/L
56 Operating Instructions, 12/2016, A5E33482052-AD
Select: 1 - Calculated
Select: 2 - Sensor Data
The program returns to P015. Calculate the
value per Parameter P690.
The program advances through parameters P691
and P692 prompting entry from the sensor name-
plate. From this data, the speed constant is calcu-
lated and automatically entered into P015.
P015-01 Speed Constant E
Pulses/m 0.0000
P691-01 Step 1: V
Drive Pully Diameter 0.00 mm
Press the following keys in the
given sequence:
For example: speed constant
of 100.3 pulses per meter
P692-01 Step 2: V
Pulses per sensor Rev. 0.00
Enter Pulses
P015-01 Speed Constant V
Pulses/m 0.0000
This value is calculated. For manual or automatic calculation, refer to P690 on page 143. To program
Differential Speed (P015-02), follow steps above for P015-01.
P016 Belt Length V
Enter Length 0.00
m
factory set length
Press the following key:
For example
belt length of
25 m
P016 Belt Length E
Enter Length 0.000
m
For example belt length of 25 m
Press the following keys:
P017 Test Load: Weight MS 1 V
Enter test load 0.00
kg/m
If P002 Test Load Reference had been set for 2-Chain, the display would read:
P017 Test Load: Chain MS 1 V
Enter test load 0.00 kg/m
or, if 3-ECal refer to ECal parameters (P693 - P699 on page 144)
P017 Test Load: ECal MS 1 V
Enter test load 0.00 kg/m
or, if P002 is set to 1-Weight, pressing enter at P017 invokes P680 for data entry.
Press the following key:
P680 Test Load: Weight MS 1 V
1-Enter Value, 2-Enter Data 0