Start Up
6.4 Load Cell Balancing
BW500 and BW500/L
64 Operating Instructions, 12/2016, A5E33482052-AD
Note
The moisture meter is ignored during calibration. If inclinometer is used, then calibration is
adjusted based on in
cline angle.
When performing a Span Calibration where the test reference is ECal (P002 = 3), the supplied
test weight or test chain must not be applied, and the conveyor must be run empty.
btain an accurate and successful calibration, ensure that the required criteria are met.
Calibration Criteria (Page 199).
With the conveyor stopped and locked out, apply the test weight or chain to the scale as
instructed in the associate manuals; then start the conveyor.
Press the following key:
Span Calibration Current Span
Setup test Press ENTER to Start
0
For example: the current span count.
Press the following key:
Initial Span Calibration in progress
%
The span count being calculated while calibration is in progress.
The duration of the Span calibration is dependent on speed (P014), length (P016), and
revolutions (P360) of the belt.
Span Count Too Low
● signal from load cell too low, ensure proper test weight or chain is applied during calibration
● check for proper load cell wiring and ensure shipping brackets have been removed.