MODULARIS Uro Plus SPL1-130.840.02 Page 4 of 8 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
5 - 4 Iso-center with X-Ray
• If not, adjust the shock wave head as shown in Fig. 6 (D or E):
- Loosen the screws (5/Fig. 4) on both sides.
• The shock wave head must be moved in parallel (see scale 7/Fig. 4).
The screw (6/Fig. 4) is also located on the other side of the shock wave head support.
If the shock wave head cannot adjusted, "rotate" the docking plate at the SIREMOBIL
Iso-C , i.e. the back part is higher than the front (8/Fig. 5).
• If the connecting position of the docking plate has changed, check the settings for the lift
stop switch on the SIREMOBIL Iso-C.
- Tighten the screws (5/Fig. 4) on both sides.
Concluding work 5
• Perform the protective conductor measurement.
• Perform a function check.
-D:Move the C-arm into the 20° position
as shown in (D/Fig. 6)
(viewed from the table).
-E:Move the C-arm into the 20° position
as shown in (E/Fig. 6)
(viewed from the table).
- Correct the position of the shock wave
head with the screw (6/Fig. 4) as
shown in (D/Fig. 6).
- Correct the position of the shock
wave head with the screw (6/Fig. 4)
as shown in (E/Fig. 6).