7 - 1
Siemens AG SPL1-130.840.02 Page 1 of 2 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Memoskop Programming 7
Standard programming 7
The SIREMOBIL Iso-C with LITHOSTAR MODULARIS is delivered with an organ program
that is programmed with one of the following default values:
The changes from Standard Organ program 1 appear in bold and are underlined.
The work procedure is described in the SIREMOBIL Iso-C user manual.
The program name is "Litho".
Organ program 1, Litho (User Setup, CTRL+U)
Mode of operation DL IDL DR SUB Roadmap
Program name Standard Standard Standard Standard Standard
Dose rate MID HIGH
HIGH HIGH HIGH
SIREMATIC normal HC2
HC2 n.a.
(DR850W)
n.a.
(Iodine)
n.a.
(Iodine)
SIREMATIC push HC2 HC2 n.a.
(DR850W)
n.a. n.a.
Noise red. Low K = 4
K = 4 K = 4 n.a. n.a.
Noise red. High K = 8
K = 8 K = 16 n.a. n.a.
Dose reduction low n.a. high n.a. n.a. n.a.
Dose reduction high n.a. high n.a. n.a. n.a.
Auto LIH disk transfer NO NO n.a. n.a. n.a.
Auto transfer each image n.a. NO n.a. n.a. n.a.
Disk transfer rate 0 n.a. n.a. n.a. n.a.
Auto disk transfer n.a. n.a. NO
n.a. n.a.
Subtraction n.a. n.a. n.a. MAX MAX
Landmark n.a. n.a. n.a. 10%
NO
Video signal at document n.a. n.a. n.a. POS POS
Image display n.a. n.a. n.a. POS POS