LITHOSTAR MODULARIS as "LithoShare" Execution 5
LithoShare
LithoShare is the name of the LITHOSTAR MODULARIS that can be coupled to different
SIREMOBIL Iso-C (they must be prepared) without having to newly reset the iso-center.
In this way, the user acquires several SIREMOBIL Iso-C but only one LITHOSTAR MODU-
LARIS.
Adjustment 5
Setting the iso-center is performed in the same manner as described in the paragraph
"LITHOSTAR MODULARIS from support arm serial number 0051".
Setting the iso-center must be performed on each SIREMOBIL Iso-C.
Adjusting the potentiometer on the angulation motor can only be performed on one
SIREMOBIL Iso-C.
The mechanical adjustment of a motor can only be performed on one SIREMOBIL
Iso-C.
A "Master SIREMOBIL Iso-C" must be defined where the potentio-
meter adjustment and motor adjustment is performed.
If parts pertaining to the iso-center are replaced in a "LithoShare" system, then the
system that received the replaced parts must always be adjusted.
i.e.
Replace shock wave head -> adjustment of iso-center on the LITHOSTAR MODULARIS
Replace switch in the SIREMOBIL Iso-C -> adjustment of iso-center on the
SIREMOBIL Iso-C.