LITHOSTAR MODULARIS 4 - 33
Siemens AG SPL1-130.840.02 Page 33 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Curved Probe 4
Loosen the 8 screws (4 are not visible) (8/Fig. 40) and remove the sleeve (7/Fig. 40).
Attach the sleeve on the new probe. Pay particular attention to the positioning, the
narrow side with the orange dot (6/Fig. 40) and (5/Fig. 40) must form a line.
Tighten the 8 screws.
Check the ultrasound iso-center.
Concluding work 4
Perform the protective conductor measurement.
Test all functions from the control panel.