EasyManua.ls Logo

Siemens MODULARIS Uro Plus - Adjusting the 20° Position (Serial 0051); Concluding Iso-center Work (Serial 0051)

Siemens MODULARIS Uro Plus
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Iso-center with X-Ray 5 - 7
Siemens AG SPL1-130.840.02 Page 7 of 8 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Adjusting the 20° position
Move the C-arm into one of the two 20° positions.
Switch FL ON briefly.
The ball must be in the center of the cross (refer to Fig. 8), if not:
- decouple the LITHOSTAR MODULARIS from the SIREMOBIL Iso-C and slightly
loosen the eight screws (1/Fig. 10).
- Recouple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C.
- Move the C-arm into the 20° position.
- Use the adjustment screws (2/Fig. 10) to adjust the support arm so that the ball is
located in the cross-hairs (refer to Fig. 8).
- Once the iso-center is correctly set, couple the LITHOSTAR MODULARIS again and
tighten the eight screws (1/Fig. 10) with a torque of 10 Nm.
- Recouple the LITHOSTAR MODULARIS to the SIREMOBIL Iso-C and check the
switch setting on the SIREMOBIL Iso-C for column stop and reset if necessary.
- Angulation motor - check adjustment.
Concluding work 5
Perform the protective conductor measurement.
Perform a function check.
Fig. 10
The position of the ball must not be identical in both 20° positi-
ons. The tolerances must however be maintained, refer to
Figure 8.
1
2
2
NOTICE

Table of Contents

Other manuals for Siemens MODULARIS Uro Plus

Related product manuals