E00 Error Text: System was switched on.
Possible
cause/action:
E99 Error Text: The system was switched off without deleting the therapy data.
Possible
cause/action:
Press the Reset button.
E10 Error Text: Cooling system losing water.
Possible
cause/action:
Is water running out of the hoses?
P
min
-switch o.k.?
Board D3.X3 connection o.k.?
Fill the cooling circuit.
E11 Error Text: Maximum pressure in the cooling circuit exceeded.
Pressure switch P
max
responded.
Possible
cause/action:
Are the hoses blocked, kinked?
P
max
-switch o.k.?
Board D3.X3 connection o.k?
Is the cooling unit operating?
Check the fill level in the cooling circuit.
E12 Error Text: Minimum pressure in the cooling circuit exceeded.
Pressure switch P
min
responded.
Possible
cause/action:
Is water running out of the hoses and hose connections?
Air in the cooling system?
Check the fill level in the cooling circuit.
NOTE
LITHOSTAR MODULARIS Error List 2