LITHOSTAR MODULARIS Error List 2 - 11
Siemens AG SPL1-130.840.02 Page 11 of 14 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Internal errors, E80 E89 2
E80 Error Text: Flash memory defective.
Possible
cause/action:
Not applicable for the user.
Read out the memory and perform a reset at the next opportu-
nity.
E81 Error Text: CMOS pulse counter incorrect.
Possible
cause/action:
If this error occurs repeatedly, board D3 must be replaced.
E82 Error Text: Pulse counter backup incorrect
Possible
cause/action: If this error occurs repeatedly, board D3 must be replaced.
E83 Error Text: CMOS chip card data incorrect
Possible
cause/action:
E85 Error Text: Measurement acquisition incorrect.
Possible
cause/action:
E86 Error Text: Flash memory defective
Possible
cause/action:
Not applicable for the user.
Read out the memory and perform a reset at the next opportu-
nity.
E89 Error Text: Watchdog responded.
Possible
cause/action:
If this error occurs repeatedly, board D3 must be replaced.
Appears each time a reset is done with switch S3 on board D3.
This error may occur if the system is switched off during initiali-
zation.