Page #53 background image
LITHOSTAR MODULARIS 4 - 13
Siemens AG SPL1-130.840.02 Page 13 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
IWAKI cooling pump 4
Removal 4
Switch the system off.
Open the system cover.
Remove the fasteners on the pump.
Remove the motor connections on the terminal block.
Loosen the screws to remove the pump.
Installation 4
Tighten the pump.
Connect the motor connections.
Connect the fasteners on the pump (The black-marked water hoses must be connected).
Fill the cooling circuit (refer to the section "Filling the cooling circuit").