IWAKI cooling pump 4
Removal 4
• Switch the system off.
• Open the system cover.
• Remove the fasteners on the pump.
• Remove the motor connections on the terminal block.
• Loosen the screws to remove the pump.
Installation 4
• Tighten the pump.
• Connect the motor connections.
• Connect the fasteners on the pump (The black-marked water hoses must be connected).
• Fill the cooling circuit (refer to the section "Filling the cooling circuit").