LITHOSTAR MODULARIS Error List 2 - 13
Siemens AG SPL1-130.840.02 Page 13 of 14 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Displays on board D3 2
Once the system is turned on, different symbols are displayed chronologically on the dis-
play of board D3.
Text on the Console 2
"Chip card invalid" 2
This message is displayed when:
- the gold card is still logged on a different LITHOSTAR MODULARIS
- board D3 was replaced and the "init counter" on the gold card is already set at 3.
- the card is defective.
In normal mode, check whether the LED 174 on board D3 blinks temporarily and then
remains lit when turning it on.
LITHOSTAR MODULARIS SIEMENS 2
This message is displayed when too many errors occur. In this case, use the service PC
to generate an error log.
Missing Iso-center cross on the ultrasound system 2
There is no communication between the ultrasound system and the LITHOSTAR
MODULARIS.
Display Possible cause/action
0 System o.k.
(only with the ultrasound localization option and if a communication was set
up to the ultrasound device.)
d Software download.
1 Data transfer, the decimal point is flashing.
4 The receiving file is being checked.
5 Software download time too long.
7 Everything o.k.
(only if no ultrasound localization option is available or if the ultrasound device
is turned off.)
- The service switch is set to normal mode - only for boards that still have no
software.