MODULARIS Uro Plus SPL1-130.840.02 Page 4 of 14 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
2 - 4 LITHOSTAR MODULARIS Error List
Controls, E26 – E29 2
E26 Error Text: ECG option not installed.
Possible
cause/action:
Warning message only.
E27 Error Text: Control unit.
Possible
cause/action:
Key sticks.
Replace control panel.
E28 Error Text: C-arm swivels in the wrong direction.
Possible
cause/action:
Cables connected wrong at the potentiometer or the motor?
Correct tension on the belt?
Check the setting of the SIREMOBIL angulation brake
(see E23).
E29 Error Text: Shock wave release button sticks.
Possible
cause/action:
Check control unit. Circuit diagram J1048 – 25.