Service software 3 - 9
Siemens AG SPL1-130.840.02 Page 9 of 18 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
select option: c 3
• Press the "c" key to display the following text::
- *1: sequential number of the card inserted.
- *2: serial number of the card.
- *3: a "gold" card was inserted, i.e. the input slot is closed.
- *4: init-counter in the above example is init=2, i.e. the gold card was logged onto two
different D3 boards.
- *5: n.a.
- *6: n.a.:
- *1:sequential number of the card inserted
- *2: serial number of the card
- *3: a "pay per use" card was inserted
- *4: shock waves still present on the card
- *5: internal system card counter
- *6: number of shock waves registered on inserting the ppu card
Text: 4: 531...0 G init=2 .... ..... 04/13 ... 418
Explan.: *1 *2 *3 *4 *5 *6 *7 *8
Card ejected - *7: Date of ejection
- *8: System shock wave counter value when card ejected
Card in system - *7: Text: "Card locked to unit"
- *8: ............
Text: 5: 531...1 P 89956 0 0 04/13 ... 418
Explan.: *1 *2 *3 *4 *5 *6 *7 *8
Card ejected - *7: Date of ejection
- *8: System shock wave counter value when card ejected
Card in system - *7: Text: "Card locked to unit"
- *8: ............
=> C
Card Data:
# Serial number Type -------< units >-------- eject-date/time pulse-count
GOLD init-cnt ... ... system
PPU on card int. from card
------------------------------------------------------------------------------
1:
2:
3:53148300000000 G init=2 .... .... 04/13/1999 15:48.11 418
4:53149100000001 P 89956 0 0 04/13/1999 16:00.21 418
5:53148300000000 G init=3 .... .... 04/13/1999 16:40.51 418
6:53149100000001 P 84956 0 5000 04/13/1999 17:45.00 418
7:
8:
9:
10:53149100000001 P 84956 5000 0 card locked to unit ......
-- end of card data --