MODULARIS Uro Plus SPL1-130.840.02 Page 8 of 18 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
3 - 8 Service software
Example:
- 1:0012 - 12 shock waves with an energy step between 0.1 and 1.0 (energy steps
between 0.1 and 1.0 are summarized in intervals of 0.1) released).
- **E60 - error 60 has occurred.
- 32:0041 - 41 shock waves with energy steps 3.2 released.
- **E60 - error 60 has occurred again.
Once all of the therapy data for a day is listed, the above text will be displayed. Press the
"d" key to delete all therapy data.
Content of Therapy log (selected day):
Patient no. 0001 start of therapy: 03/03/1999 17:03.15
---------- table of energy-level (*10) : pulses -----------
1:0012 ** E60: 32:0041 ** E60: -:.... -:.... -:.... -:.... -:.... -:....
-:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:....
-:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:....
-:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:.... -:....
End of therapy: 03/03/1999 17:03.24
---- end of therapy data ----
press key 'D' to delete all therapy data, any other key to abort
please save therapy data to file before deleting it !