LITHOSTAR MODULARIS 4 - 23
Siemens AG SPL1-130.840.02 Page 23 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Angulation drive (C-arm drive) 4
Open the cover of the angulation drive.
Disconnect the motor cable coming from the unit at the clamp (K/Fig. 26).
Desolder the cables at the potentiometer (P/Fig. 26); when doing this, record the color
coding of the cables to the potentiometer connections to ensure correct reconnection.
Remove the old angulation drive completely.
Install the new angulation drive. Do not tighten the mounting screws yet (5/5a/Fig. 26).
Adjust the C-arm to 0°.
Couple the LITHOSTAR MODULARIS to the SIREMOBIL Iso -C.
- The pin (6/Fig. 27) should go easily into the guide; if not, rotate the motor via the control
panel of the LITHOSTAR MODULARIS.
- The screws can still be loosened (5/5a/Fig. 26). Follow the correct sequence when
tightening (5 up, 5 down, 5a up, 5a down/ Fig. 26).
Set the potentiometer to 500 ± 10 ohms or 5.0 ± 0.1 volts (center position).
The potentiometer setting can also be checked with the Service PV (Chapter 3).
Tighten the toothed belt.
Adjust the C-arm according to the description on pages 3 - 10.
Reattach the motor cover.

Table of Contents

Other manuals for Siemens MODULARIS Uro Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MODULARIS Uro Plus and is the answer not in the manual?

Siemens MODULARIS Uro Plus Specifications

General IconGeneral
BrandSiemens
ModelMODULARIS Uro Plus
CategoryIndustrial Equipment
LanguageEnglish

Related product manuals