EasyManua.ls Logo

Siemens MODULARIS Uro Plus - Check of Image Tilt (Sonoline G20)

Siemens MODULARIS Uro Plus
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Iso-center with ultrasound 6 - 3
Siemens AG SPL1-130.840.02 Page 3 of 16 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Check of image tilt 6
Check that the isocenter phantom (1/Fig. 5) is filled with water, if not remove the screw
(2/Fig. 5) and fill water with a syringe (3/Fig. 5).
No air bubbles may be present. Screw the screw back in.
Hinge the isocenter phantom up.
Fill the water bladder with water (approx. 100 - 120 ml, i.e. 5 - 6 syringes a 20 ml)
(1/Fig. 6):
- Fill a syringe with distilled water. Remove the air remaining in the syringe.
- Push the syringe needle through one of the valves (2/Fig. 6) and empty the contents of
the syringe into the water bladder. The valve has a self-closing membrane.
- Repeat the filling process until the water bladder if filled with approx. 200 - 120 ml
water.
- After filling there must not be any air bubbles in the water bladder; withdraw air bubbles
with the syringe.
Apply contact gel to the isocenter phantom.
Fasten the water bladder (5/Fig. 1) with the holding part (6/Fig. 1) to the isocenter
phantom.
Apply contact gel to the ultrasound probe.
With the "Image field/zoom" knob set the value "50" (Fig. 7) - bottom left in the ultrasound
image.
Push the probe so far forwards (3/Fig. 1) until the isocenter cross is approximately in the
center of the SONOLINE Prima screen.
Press the menu key and display this.
Fig. 5 Fig. 6
The isocenter must be checked mit every type of probe and after
every
probe change
2
1
3
2
1
NOTE

Table of Contents

Other manuals for Siemens MODULARIS Uro Plus

Related product manuals