LITHOSTAR MODULARIS 4 - 7
Siemens AG SPL1-130.840.02 Page 7 of 34 MODULARIS Uro Plus
Medical Solutions Rev. 03 04.05 CS PS 24
Replace isocenter phantom 4
System con-
figuration
n = several
1 LITHOSTAR MODULARIS
and
1 SIREMOBIL Iso-C
1 LITHOSTAR MODULARIS
and
n SIREMOBIL Iso-C
Replace iso-
center phan-
tom
To support arm S0050
Adjustment of iso-center on the shock
wave head; refer to chapter "Iso-cen-
ter with X-ray - to support arm serial
number 0050".
It is important to follow this workflow to
avoid having to readjust the iso-center on all
SIREMOBILE!
Prior to replacing the iso-center phantom,
view the iso-center (the three C-arm
positions) and if possible, save it or make a
hardcopy.
Replace the iso-center phantom.
Check iso-center; the setting must be
identical to that of the defective iso-center
phantom.
If an adjustment is necessary, it must be made
via the shock wave head setting; refer to
chapter "Iso-center with X-ray - to support
arm serial number 0050".
From support arm S0051
Adjustment of iso-center on the
SIREMOBIL Iso-C; refer to chapter
"Iso-center with X-ray - from support
arm serial number 0051".