MODULARIS Uro Plus SPL1-130.840.02 Page 12 of 34 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
4 - 12 LITHOSTAR MODULARIS
Hose pump head in the coupling circuit 4
Removing the hose pump head 4
• Remove the system cover and the cover of the shock wave head.
• Empty the coupling circuit (refer to the section "Emptying the coupling circuit").
• Switch the system off.
• Remove the safety latch on the hose pump head.
• Press the black lock (1/Fig. 10/11) upward or downward and remove the hose pump
head from the drive.
• Open the 2 hose clamps and remove the hoses.
• Remove the hose pump head.
Installing the hose pump head 4
• Place the hose pump head on the drive shaft and lock into place.
• Reconnect the 2 hoses with the hose clamps.
Ensure that the hoses are the correct length, i.e. cut them to the same length as the old
hoses.
• Fill the coupling circuit (refer to "Filling the coupling circuit").
Hose pump head in the cooling circuit 4
Removing the hose pump head 4
• Switch the system off.
• Open the system cover.
• Empty the cooling circuit (refer to the section "Emptying the cooling circuit").
• Remove the safety latch on the hose pump head.
• Press the black lock (1/Fig. 10) downward or upward and remove the hose pump head
from the drive.
• Open the 2 hose clamps and disconnect the hoses.
• Remove the hose pump head.
Installing the hose pump head 4
• Place the hose pump head on the drive shaft and lock into position.
• Reconnect the 2 hoses with the hose clamps.
Ensure that the hoses are the correct length, i.e. cut them to the same length as the old
hoses.
• Fill the cooling circuit (refer to the section "Filling the cooling circuit").