MODULARIS Uro Plus SPL1-130.840.02 Page 24 of 34 Siemens AG
Rev. 03 04.05 CS PS 24 Medical Solutions
4 - 24 LITHOSTAR MODULARIS
SIREMOBIL Iso-C switch
Fig. 28 Fig. 29
System con-
figuration
n = several
1 LITHOSTAR MODULARIS
and
1 SIREMOBIL Iso-C
1 LITHOSTAR MODULARIS
and
n SIREMOBIL Iso-C
Remove doc-
cking plate
To support arm S0050
Adjustment of iso-center on the shock
wave head; refer to chapter "Iso-cen-
ter with X-ray - to support arm serial
number 0050."
It is important to follow this workflow to
avoid having to readjust the iso-center on all
SIREMOBILE!
• Prior to removing the docking plate, view the
iso-center (the three C-arm positions) and if
possible, save it or make a hardcopy.
• Remove the docking plate (mark the position
on the SIREMOBIL Iso-C).
• Perform the necessary work on the
SIREMOBIL Iso-C.
• Tighten the docking plate back into place.
• Check iso-center; the setting must be
identical to how it was prior to removing the
docking plate.
If an adjustment is necessary, it must be made
via the iso-center setting on the SIREMOBIL
Iso-C; refer to chapter "Iso-center with X-ray -
form support arm serial number 0051."
From support arm S0051
Adjustment of iso-center on the
SIREMOBIL Iso-C; refer to chapter
"Iso-center with X-ray - from support
arm serial number 0051."
Damage of docking plate.
The lift switch-off in the SIREMOBIL Iso-C may not switch off if the
floor is not level. If this is the case, make use of the additional
steering castors (see also Operating Instructions).
7
S1
S2
NOTICE