Measuring ranges for resistance thermometers Pt 100 climate
Table 14-16 SIMATIC S7 format: Measuring ranges Pt 100 climate in °C and °F
Pt 100 cli‐
mate in °C
(1 digit =
0.01 °C)
Units Pt 100 cli‐
mate in °F
(1 digit =
0.01 °F)
Units Range
Decimal Hexa‐
decimal
Decimal Hexa‐
decimal
> 155.00 32767 7FFF
H
> 311.00 32767 7FFF
H
Overow
155.00
:
130.01
15500
:
13001
3C8C
H
:
32C9
H
311.00
:
266.01
31100
:
26601
797C
H
:
67E9
H
Overrange
130.00
:
-120.00
13000
:
-12000
32C8
H
:
D120
H
266.00
:
-184.00
26600
:
-18400
67E8
H
:
B820
H
Nominal range
-120.01
:
-145.00
-12001
:
-14500
D11F
H
:
C75C
H
-184.01
:
-229.00
-18401
:
-22900
B81F
H
:
A68C
H
Underrange
< -145.00 -32768 8000
H
< -229.00 -32768 8000
H
Underow
Measuring ranges for resistance thermometers Ni 100 standard
Table 14-17 SIMATIC S7 format: Measuring ranges Ni 100 standard in °C and °F
Ni 100
standard in
°C (1 digit =
0.1 °C)
Units Ni 100
standard in
°F (1 digit =
0.1 °F)
Units Range
Decimal Hexa‐
decimal
Decimal Hexa‐
decimal
> 295.0 32767 7FFF
H
> 563.0 32767 7FFF
H
Overow
295.0 2950 B86
H
563.0 5630 15FE
H
Overrange
: : : : : :
250.1 2501 9C5
H
482.1 4821 12D5
H
250.0 2500 9C4
H
482.0 4820 12D4
H
: : : : : : Nominal range
-60.0 -600 FDA8
H
-76.0 -760 FD08
H
-60.1 -601 FDA7
H
-76.1 -761 FD07
H
Underrange
: : : : : :
-105.0 -1050 FBE6
H
-157.0 -1570 F9DE
H
< -105.0 -32768 8000
H
< -157.0 -32768 8000
H
Underow
Analog electronic modules
14.8 Representation of analog values
ET 200iSP
380 Operating Instructions, 11/2022, A5E00247483-AK