Replacing boards / replacing components 6 - 37
Siemens AG SPR2-230.061.01 Page 37 of 48 SIREMOBIL Iso-C
Medical Solutions Rev. 13 12.04 CS PS 24 System Manual
Cutt-off delay - PC board D210
• In the monitor trolley, PC board D210 is installed above the mains filter (Fig. 11) .
• Disconnect the system from the power supply by pulling the mains plug out of the mains
socket.
• Loosen the 4 fastening nuts of PC board D210.
• Unplug all connectors.
• Replace board D210 and fasten the new PC board with the 4 nuts.
• Reconnect all connectors.
• Connect the monitor trolley to the power supply again.
• Switch the monitor trolley on and wait for the run-up.
• Check the cut-off delay.
• After the power OFF button has been pressed, the monitor trolley must switch off after
approx. 3 minutes.
Fig. 11
Electrical voltage!
See Chapter 1, Safety Information.
Switch system power supply off and disconnect the power plug.
Only applies to SIREMOBIL Iso-C 3D.
WARNING
NOTE