EasyManua.ls Logo

Silca UC 199 - Calibrating the OPTIONAL Automatic Clamp L1 (2 Nd Calibrating)

Silca UC 199
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustments for X and Y can be made manually.
Press SHIFT+ENTER to see on the display:
L100-STANDARD S: C
Manual adjustment:
X=+000 Y=+000
Adjustments for X and Y can be made manually.
Press SHIFT+ENTER to see on the display:
L100-STANDARD S: D
Manual adjustment:
X=+000 Y=+000
NOTE: (*) Note: if you are conrming a measurement with a deviation of over +/-30 hundredths of a
millimetre the display will show:
ATTENTION
X/Y gauged beyond
the advisable
tolerance range!
Repeat the gauging procedure.
5.7.1.2 CALIBRATING THE OPTIONAL AUTOMATIC CLAMP L1 (2ND CALIBRATING)
ATTENTION: for possessors of the optional feeder unit and only after having calibrated the
standard clamp (1st calibration)
- CUTTER SIDE CLAMP-
Beforestartingthecalmpcalibration,checkthattherearenokeys(orother)ttedtoclamps.
NOTE: check that the loader and automatic clamp are properly tted (see Ch. 5.8.3).
Select L1-AUTOMATIC and press ENTER; the display will show:
L1-AUTOMATIC S: A
Manual adjustment:
X=+000 Y=+000
To calibrate [START]
NOTE: the Automatic clamp must have two jaws A.
Press START, the display will show:
Fit key blank
on loader
See operating manual.
[START]
-
raise the safety shield.
-
tatleast1testkeyblankZ3intotheloader,withitsbacktowardstheoperator.
-
close the safety shield.
-
pressSTARTtoseeonthedisplay,rst:
Operating Manual UC199
Copyright Silca 2012
47

Table of Contents

Related product manuals