1-4
056404-443 Operating Manual Safety Precautions
WARNING – DISPOSAL HAZARD:
WARNING – INFECTION HAZARD:
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
During Backup Control, tabletop functions are available even if table floor locks are NOT engaged.
LOCK table before moving tabletop or transferring patient.
Use of accessories, transducers or cables with this table, other than those specified, may result in
increased emissions or decreased immunity of the table.
Do not store items on table base. Doing so may result in equipment damage or inadvertent tabletop
movement placing the patient and/or user at risk of injury.
Failure to keep all personnel and equipment clear of the table before actuating any inertia-driven or
power-driven movement could result in table damage and/or personal injury.
Hand Control must be attached to table at all times to enable safety stop and all modes of operation.
With patient in normal or reverse orientation, table in Trendelenburg and slide at head limit, center
tabletop before attempting to raise table to level or move into Reverse Trendelenburg.
Read and understand all instructions presented in this section before using the Universal Hand Control.
Follow each step in the order presented in these instructions. If you need technical assistance or
additional instructions, please contact STERIS.
Dual-articulating headrest side rails support 100 ft-lbs. (135 N•m) maximum. The headrest end rail
supports 50 ft-lbs. (67 N•m) maximum. Exceeding these limits could result in injury or equipment
damage.
This product contains materials which may require disposal through appropriately licensed and
permitted hazardous waste management firms.
To protect against aerosols being reflected from potentially contaminated surfaces, wear rubber or
plastic gloves, masks and eye protection, and follow OSHA blood-borne pathogens standards when
cleaning.
When cleaning/disinfecting table, do not use alcohol as it does not have sufficient cleaning/disinfection
properties.
When moving the table to or from point of use, roll it carefully at moderate speed and only over smooth
floors. Maximum floor clearance is 1/4" (6.4 mm). Avoid door and elevator jambs, and obstructions
greater than 1/4" (6.4 mm). If necessary, lift table over obstructions, onto trucks, etc. Lift table evenly
and only by the table base. DO NOT transport articles (including accessories) on top of the table, and
DO NOT use a forklift to move the uncrated table.
Hang the hand control from side rail of the table when not in use to avoid possible damage to the
control.
When cleaning/disinfecting table, thoroughly read the cleaning fluid directions for use and follow all
directions and cautions as shown.
Do not spray cleaning fluid into electric receptacles and avoid spraying directly on emergency backup
buttons or into clearance space. Spray or drippage may settle onto electric circuits inside table causing
corrosion and loss of function.