7-1
Troubleshooting Operator Manual 056404-443
This section of the manual describes the types of table malfunctions which may occur and indicates probable
causes.
If you are unable to correct the situation following the Troubleshooting Charts (or if a problem occurs which is not
described on the charts) or S
ECTION 6, ROUTINE MAINTENANCE of this publication, please contact STERIS. Service
charges may be incurred, consult your warranty for details. A trained service technician will promptly place your
STERIS 4095 General Surgical Table in proper working order.
NOTE: Never permit unqualified persons to service the table. If you are unable to correct a table malfunction,
please contact STERIS.
Calling For Service: First, try to define the situation and determine whether you can solve it yourself (see
Table 7-1 or Table 7-2). If you cannot solve the situation, call STERIS and give the following information for the
table: model number, serial number, date of purchase.
NOTE: To prevent possible hand control functioning issues, return table to level between a series of hand control
commands.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
• Failure to perform periodic inspections of the table could result in serious personal injury or
table damage.
• Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained
service personnel. Maintenance performed by unqualified personnel or installation of
unauthorized parts could cause personal injury, result in improper equipment performance,
invalidate the warranty, or result in costly damage. Contact STERIS regarding service options.
CAUTION – POSSIBLE EQUIPMENT DAMAGE: When moving the table to or from the point of use,
roll it carefully at moderate speed and only over smooth floors. Maximum floor clearance is
1/4" (6.4 mm). Avoid door and elevator jambs, and obstructions greater than 1/4" (6.4 mm). If
necessary, lift table over obstructions, onto trucks, etc. Lift table evenly and only by the table
base. DO NOT transport articles (including accessories) on top of the table, and DO NOT use a
forklift to move uncrated table.
Table 7-1. Troubleshooting Chart
SITUATION ACTION
Hand Control does not activate when table is plugged
into electrical outlet and any pushbutton (except
STOP) is pressed .
Main battery fuse blown – replace.
Faulty power cord – replace.
Hand control not plugged in properly – correct.
Faulty hand control** – replace.
Call STERIS.*
Table remains on battery power even when main
power cord is plugged into facility power.
Incorrect power cord connection – correct.
Faulty power cord – replace.
Main fuse blown – replace.
Call STERIS.*
TROUBLESHOOTING
7