5-13
Operating Instructions Operator Manual 056404-443
LEG UP: When depressed and pumping simultaneously, the tabletop leg section raises.
LEG DN: When depressed and pumping simultaneously, the tabletop leg section lowers.
KIDNEY UP: When depressed and pumping simultaneously, the tabletop KIDNEY Elevator raises to the
maximum 4" (110 mm).
KIDNEY DN: When depressed and pumping, the tabletop Kidney Elevator Mechanism lowers.
FLOOR LOCK: When depressed and pumping simultaneously, floor locks are slowly lowered. To ensure table is
correctly secured to the floor, gently push on table.
FLOOR UNLOCK: When depressed, floor locks are slowly released. To ensure floor locks are completely
disengaged, gently push on table.
NOTE: When table is in the Backup FLOOR UNLOCK configuration, tabletop functioning IS active. FLOOR LOCK
table before moving tabletop.
STOP: When depressed, all hand control function buttons deactivate. All functions are stopped immediately.
5.9.3 Return Backup Control System
When situation has been corrected and normal table operation is available, replace Backup Control System as
follows (see Figure 5-4):
1. Turn Backup Hand Control face down.
2. Wind cable and place under controller.
3. Slide controller and cord back into table base.
4. Fold pedal back to original position.
Figure 5-5. Backup (Override) Hand Control
TREND
REVERSE
TREND
HEIGHT UP
HEIGHT DN
BACK UP
BACK DN
LEG UP
LEG DN
FLOOR
LOCK
STOP
CAUTION
Back-up mode
Auto Limit Sensors NOT available
USE WITH FOOT PUMP
SLIDE HEAD SLIDE FOOT
KIDNEY UP
KIDNEY DN
FLOOR
UNLOCK
TILT LEFT
TILT RIGHT